Flow sensor comparison

on flexible tubes in realistic test environment

SONOFLOW C0.55 Clamp-On Sensors - Non-Invasive | Contamination-freeSONOFLOW C0.55 Clamp-On Sensors - Highly accurate & PreciseSONOFLOW C0.55 Clamp-On Sensors - For multiple tubing sizes & materials
SONOFLOW C0.55 Clamp-On Sensors - Non-Invasive | Contamination-freeSONOFLOW C0.55 Clamp-On Sensors - Highly accurate & PreciseSONOFLOW C0.55 Clamp-On Sensors - For multiple tubing sizes & materials

AT A GLANCE

  • Ultrasonic flow sensors compared to turbine meters
  • Definition of accuracy and precision
  • SONOFLOW CO.55 V2.0 - convinces with highest precision and accuracy
  • SONOFLOW CO.55 V2.0 - contamination-free clamp-on system for use in microbiologically critical environments

Company Althea, one of the leading North American contract manufacturers specialized in cGMP manufacturing, analytical development, aseptic filling and protein delivery technologies was looking to evaluate flow systems for multiple applications throughout their facilities. One of the main issues was to replace scales for volume measurement by flow sensors. For that reason, three competing flow sensor set-ups were tested for their measuring performance and suitability for the customers’ specific applicative needs.

Content of the test

Test Setup

SensorsSONOFLOW CO.55/160 V2.0 / Competitor X / Competitor Y
TubingRM-2708, 0.5” ID, 0.75” OD
PumpWatson Marlow 620S pump, ALT-34345
Data LoggerOmega Data Logger, OM-CP-QUADPROCESS-25MA
Feeding100L plastic carboy
Constraints Water at room temperature

Sensor Specifications

SONOTEC
(ultrasonic clamp-on flow sensor)
Competitor X
(disposable turbine sensor)
Competitor Y
(ultrasonic clamp-on flow sensor)
Accuracy±1 %±5 %±3 %
Precission±0.5 %≤1 %±20 mL/min (max)
Resolution   0.001 ml/min1 ml/min2 ml/min

Difference Accuracy and Precision

There is often confusion between accuracy and precision, especially when someone asks how accurate SONOTEC sensors actually are? Accuracy describes the error between the real and the measured value. It is an indication about how close you can get to the correct value. Reversely, precision describes the random spread of measured values around the average measured values. It finally defines how consistent you can get the results. The standard deviation (STDEV) of a measuring sequence stands for a measure of precision.

Accuracy and Precision

Precision is by far the most critical determinant for any measuring device and the basis for an accurate measurement. A precise flow sensor can be calibrated to become an accurate one; the reverse case is not possible.

Resolution indicates the smallest detectable change in flow that a sensor can measure.

Accuracy and Precision – A Conceptual Delimitation (Illustration Example with fictitious values)


Flow Sensor Comparison on Flexible Tubes – Sensor Setup

Test set-ups A/B/C/D/E/F represent a 60 minutes run each, while the sensors have been re-clamped from test to test as visualized in the chart above. Setup Z, however, was the only 180 minutes run performed.

 

 

The sensor setup was chosen to minimize systemic measurement errors while comparing the sensors to each other. All test conditions were kept as stable as possible by using the same sensor and application set-up in steady ambient conditions such as temperature and air pressure. All sensors have been calibrated to the applied test setup to finally refer to the pump output at 3.5 litres per minute.

Flow Sensor Comparison on Flexible Tubes – Measuring Results (Pump Output: 3.5 Litres/Minute; 60/180 Minutes Runs)

Flow Sensor Comparison on Flexible Tubes – Visualized Measuring Results (Pump Output: 3.5 Litres/Minute; 60/180 Minutes Runs)

SONOFLOW CO.55 V2.0 stands out for high precision and best accuracy – Interpretation of the results

Summarizing all six 60 minutes runs and the 180 minutes run, the average standard deviation of the SONOFLOW CO.55 V2.0 sensor of 0.0406 litres per minute indicates the highest precision of all three tested sensors. This number, however, represents an average precision of 1.16 % of the SONOTEC device, compared to 4.04 % (Sensor X) and 4.82 % (Sensor Y).

As the precision specifies the repeatability of the sensor describing how consistent results could be reached; accuracy in turn, defines the error between the real and the measured value. Accuracy, indeed, is finally an issue of a high precision and an optimum calibration of the sensor.

With reaching the best values of the standard deviation during each run, the SONOFLOW CO.55 V2.0 shows the best precision of all competing sensor set-ups. This outstanding performance makes sure to also reach consistently high accuracies, ranking the SONOFLOW CO.55 V2.0 top in five of seven of the simulations realized. This, however, could be even topped by an improved calibration to the test environment.      

The high resolution of the SONOFLOW CO.55 V2.0 is thus also one of the key reasons reaching these exceptional results.

SONOFLOW CO.55 V2.0 – Advantages at a glance

  • Non-invasive sensor solution to be easily clamped on the flexible tubing
  • No additional electronics board needed; less installation effort and less cabling
  • Available in aluminum and stainless steel housing to fulfil high GMP standards
  • Sensors are able to switch devices on its own (e.g. pump)
  • Software for specific adjustments, e.g. in-house calibration
  • Manifold outputs: 4-20 mA, RS485, 0-20 kHz, switching
  • Fast return on investment (ROI) compared to yearly costs of disposable sensors
  • Sustainable as the sensor is re-usable

Further product information

[Translate to English:] Flow Sensor SONOFLOW CO.55

SONOFLOW CO.55 | Clamp-On Sensor
...for non-invasive flow measurement at flexible tubes

Contact us for your queries.

Contact Form

By submitting the form I agree to the privacy policy.
* Mandatory Fields

Application Notes &
White Paper


Worldwide

Mandy Reber

Do you have any questions?

Mandy Reber

+49 (0)345 / 133 17-832
m.reber@sonotec.de
Print PDF-Version
×